15
000000Ć1150Ć839 IOLMaster 11.02.2004
Overview of buttons and shortcut keys
Button
Key
Function
Remarks
<
N
>
Activates data entry
screen.
For inputing new patients
as well as recalling patiĆ
ent data within the last
5 to 100 days
Input of patient data is
compulsory.
<
O
>
Activates overview mode
and surrounding field
illumination.
For coarse alignment prior
to measurement
<
A
>
Activates axial length
mode.
<
K
>
Activates corneal radius
measurement mode.
<
D
>
Activates anterior chamĆ
ber depth mode.
<
W
>
Activates WTW
determination (optional)
WTW = White To White
<
I
>
Activates IOL calculation. Calculation possible after
measurement of one eye.
After corneal refractive
surgery, it is necessary to
determine the corneal
refraction.
<
P
>
Prints results
<
S
>
Exports data.
Requires connection of an
appropriate computer to
the serial port. *
<
X
>
Sends data to diskette.
Requires diskette to be
inserted in disk drive.
<
E
>
Exits IOLMaster software
and Windows.
In case of problems: Turn
off power switch and
disconnect power cable!
*
If connecting an external PC, the
owner/operator must ensure to
meet the safety requirements as
per EN 60601Ć1Ć1 (medical
electrical systems)!
Operation