000000Ć1150Ć839 IOLMaster 11.02.2004
Notes on installation and use
q
Do not operate the instrument
Ć in explosionĆrisk areas
Ć in the presence of flammable anesthetics and volatile solvents such
as alcohol, benzine or similar chemicals.
q
Do not store or use this instrument in damp rooms. Avoid dripping
or splashing fluids near the instrument.
q
This instrument may only be modified and repaired by service
technicians employed with or authorized by Carl Zeiss Meditec.
q
The manufacturer is not liable for damages caused by unauthorized
persons tampering with the instrument; such tampering will also void
any rights to claim under warranty.
q
Use this instrument only with accessory equipment delivered by Carl
Zeiss Meditec.
q
The instrument may be operated only by instructed and trained
personnel.
q
Keep the users manuals always at hand.
q
The instrument may only be opened by service technicians employed
with or authorized by Carl Zeiss Meditec.
q
Use printers recommended by Carl Zeiss Meditec only.
q
Place the printer at a distance of at least 1.5 m from the patient's seat
at the instrument.
q
If you use a printer of Protection Class II (without protective earth
terminal), you must make sure that a power isolation transformer is
included in the power line of the printer.
q
If you use a printer of Protection Class I (with protective earth termiĆ
nal), you must make sure that a power isolation transformer is incluĆ
ded in the power line of the printer or it is connected to a separate
stationary power outlet of the room's electric installation.
q
You can obtain the required isolation transformer via our sales orgaĆ
nization.
q
Do not use a cellular telephone and other devices not complying with
the requirements of EMC Class B because signals transmitted may
cause the equipment to malfunction. The effects of radio signals on
medical instruments depend on different factors and are, therefore,
not foreseeable. To avoid electromagnetic disturbances, the instruĆ
ment has to be installed and set into operation in accordance with
the Technical Manual and using exclusively the components deĆ
livered by Carl Zeiss Meditec.
q
Additional movable multiple sockets or extension lines must not be
connected to the system.
q
For calculating the lens power with the IOL Master, do not use the
constants of the manufacturer, but constants optimized for the
optical biometry only (see also p. 42 and p. 60).