2.2.3
Changing the DIC prism in the
universal condenser
•
To loosen the cover on the condenser underside
(2-25/
), loosen both grub screws using the
SW
2
screwdriver and remove the cover.
Position the turret disk containing the DIC prism
to be exchanged in the exchange opening and
hold it on the knurled ring.
•
Unscrew retainer ring using the so-called
double-function tool (2-25/
).
•
Now screw the tool in the prism mount (2-25/
4
) and pull out DIC
).
•
2
) from the tool and
screw on the new, required DIC prism.
•
Installation of the DIC prism is made in reverse
order. Make sure that the DIC prism is oriented
correctly (groove 2-25/
5
must engage in the
pinion of the mount). Make sure that the
knurled ring of the turret disk is labelled
correctly.
2.2.4
Attachment of filter mount
•
5
) parallel to the
underside of the condenser carrier (2-26/
3
) and
4
) of the filter mount
) into the front threaded hole on the left
below the condenser carrier (2-26/
3
) until stop
using the SW 2 angled Allen key (2-26/
).
•
1
) to screw stop bolt
) into the rear threaded hole of the
condenser carrier until stop.
Fig. 2-25
Changing the DIC prism
Fig. 2-26
Attachment of filter mount