START-UP
Axioskop
2
plus
Carl Zeiss
Initial start-up
Axioskop 2
mot plus
2-12
B 40-075 e 10/03
☞
and
5
) should not be changed during coarse adjustment of the
focal spots because they influence the reflector setting in the HBO 103 illuminator.
If required, the above settings for coarse adjustment of the HBO 103 illuminator can be
performed repeatedly.
Fine adjustment using adjusting aid
•
Attach HBO 103 illuminator to microscope
stand as described in section 2.1.11 (2).
•
2
) out of the incident-
light tube (2-15/
1
). The brighter focal spot of
the HBO 103 W/2 lamp and its slightly darker
reflector image become visible in the black glass
window of the adjusting aid.
•
6
the brighter focal spot via collector adjustment.
•
Apply SW 3 ball-headed screwdriver on
adjusting screw (2-16/
4
) to focus on the slightly
darker, reflected focal spot of the lamp.
Fig. 2-17
Focal spots of HBO 103 before
coarse adjustment
Fig. 2-18
Focal spots of HBO 103 after coarse
adjustment
Fig. 2-19
Focal spot imaging via adjusting aid
and knurled knob for focusing