START-UP
Axioskop
2
plus
Carl Zeiss
Attachments and conversions
Axioskop 2
mot plus
2-20
B 40-075 e 10/03
After expiry of the average operation time of
300 h, the HBO 103 W/2 mercury vapor short-arc
lamp must be exchanged.
The operating time of the lamp can be checked on
the counter (2-30/
1
) of the HBO
100
W
transformer.
Follow the following steps when changing the
HBO 103 W/2 lamp:
•
Remove or unscrew the line plug and the HBO
103 plug from the HBO 100 W transformer.
•
Wear protective mask and safety gloves.
•
Use SW 3 ball-headed screwdriver to loosen
clamping screw (2-31/
) and remove HBO 103
) from the incident-light tube
) of the microscope stand.
•
) to bring collector to
the position at the very front (in the direction of
the light).
•
Use ball-headed screwdriver to loosen clamping
screw for lamp housing (2-32/
2
) and carefully
1
) from the lamp
) in an upward direction.
•
Press down the spring clip (2-33/
4
)
remove the
defective lamp (or transport locking pin) with
heat sink from the holder in the illuminator
socket (2-33/
2
). There is no need to break the
wire connection.
Fig. 2-30
HBO 100 W transformer
1
0
2
0
30
40
50
6
0
7
0
80
90
10
0
1
0
2
0
30
40
50
6
0
7
0
80
90
10
0
Fig. 2-31
Removing the HBO 103 illuminator