Axioskop 2
plus
START-UP
Axioskop 2
mot plus
Initial start-up
Carl Zeiss
B 40-075 e 10/03
2-11
•
Insert multi-pin plug of the HBO 103 illuminator
(2-13/
2
) of the HBO 100 W
transformer and secure it with captive coupling
ring.
•
Connect line cable to socket (2-14/
1
) of the
HBO 100 W transformer first, then connect it to
the line.
(3)
Adjustment of HBO 103
Adjustment of the HBO 103 illuminator described
below is performed in two steps: coarse
adjustment and subsequent fine adjustment. The
latter can be performed with or without adjusting
aid (2-15/
2
) on the incident-light tube (2-15/
).
Coarse adjustment:
•
Remove HBO 103 illuminator from the
microscope stand by loosening clamping screw
(2-13/
).
•
Switch on HBO 103 on HBO 100 W transformer
(2-10/
) and allow it to heat to operating
temperature. Make sure not to look into the
light exit opening of the illuminator.
•
Direct light beam against a projection area
(wall) approx. 3 m away.
•
6
) to focus on the
brighter focal spot via collector adjustment.
•
4
) and SW 3 ball-
headed screwdriver to locate and focus on the
slightly darker, reflected focal spot.
•
1
and
3
) and SW 3
ball-headed screwdriver to set the height and
side of the brighter focal spot in the same way
as the focal spot image (2-18).
Fig. 2-14
HBO 100 W transformer
Fig. 2-15
Adjusting aid
Fig. 2-16
Adjusting HBO 103