START-UP
Axioskop
2
plus
Carl Zeiss
Initial start-up
Axioskop 2
mot plus
2-10
B 40-075 e 10/03
b) Fine
adjustment
•
Attach illuminator to the stand again and tighten clamping screw (2-11/
).
•
Remove diffusion disk (3-1/
15
) and filters (3-1/
18
), if switched in, from the beam path.
•
Focus on the specimen using objective
≤
40x and search for a free object area.
•
Remove eyepiece and center lamp coil and its reflector image in the pupil image using adjusting
screws (2-12/
and
5
).
•
) to optimize the homogeneous illumination of the pupil image.
•
Switch in diffusion disk and required filters.
2.1.11
HBO 103 illuminator
(1)
Attach HBO 103 W/2 mercury vapor
short-arc lamp
The HBO 103 illuminator and the HBO 103 W/2
mercury vapor short-arc lamp are supplied in
separate packages for safety reasons.
Therefore, insertion of the HBO 103 W/2 lamp into
the lamp housing is the first step to the start-up of
this illuminator.
Insertion of a new HBO 103 W/2 lamp is described
in detail on page 2-19, section 2.2.8 entitled
"Changing the HBO 103 W/2 mercury vapor short-
arc lamp".
(2)
Attachment of HBO 103 illuminator
•
Remove cover from the incident-light tube
).
•
Insert dovetail of lamp housing in incident-light
tube (2-13/
1
) on the instrument rear and use
SW
3 ball-headed screwdriver to tighten
).
1
0
2
0
30
40
50
6
0
7
0
80
90
10
0
1
0
2
0
30
40
50
6
0
7
0
80
90
10
0
Fig. 2-13
Attachment of HBO 103 illuminator