Axioskop 2
plus
START-UP
Axioskop 2
mot plus
Initial start-up
Carl Zeiss
B 40-075 e 10/03
2-9
2.1.10
HAL 100 halogen illuminator
(1) Attachment of HAL
100 halogen
illuminator
•
Remove cover from transmitted-light tube
).
•
Insert dovetail 2-12/
)
in transmitted-light tube (2-11/
3
) on the
instrument rear and use the SW 3 ball-headed
screwdriver to tighten clamping screw (2-11/
).
•
Connect 3-pin lamp plug (2-11/
5
) to 3-pin
12 V/100 W socket (2-11/
4
) on the instrument
rear.
(2) Adjustment of HAL
100 halogen
illuminator
a) Coarse
adjustment
•
2
) and remove the
halogen illuminator from the microscope stand.
•
Switch on microscope as described under 0.
•
Direct the light beam against a projection area
(wall) approx. 3 m away.
Do not look into the light exit opening
of the illuminator.
•
3
) using SW 3 ball-
headed screwdriver in such a way that both
images of the lamp coil are as sharply focused
as possible on the projection screen.
•
Then set adjusting screws (2-12/
and
5
) in such
a way that the lamp coil of one image exactly
fills the gaps in the reflector image
).
Fig. 2-11
Attachment of HAL 100 halogen
illuminator
Fig. 2-12
Adjustment of HAL 100 halogen
illuminator