CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Axioskop 2
plus
Carl Zeiss
Troubleshooting
Axioskop 2
mot plus
4-4
B 40-075 e 10/03
4.3 Troubleshooting
Problem Cause
Remedy
Shadows or inhomogeneous
image brightness in the field of
view; the field is not entirely
The vis/fot pushrod on the camera
tube is not in the correct
(intermediate) position.
Move the vis/fot pushrod on the
phototube in the correct position
(end position), see page 3-15.
visible.
Nosepiece with objective not in
click-stop position.
Switch nosepiece with objective in
click-stop position.
Condenser not set correctly.
Set condenser correctly
(adjustment, centering),
see p. 3-24 ff.
Aperture diaphragm not set
correctly.
Set aperture diaphragm correctly
(centering, opening),
see p. 3-24 ff.
Luminous-field diaphragm not set
correctly.
Set luminous-field diaphragm
correctly (centering, opening),
see p. 3-24 ff.
Filter not inserted correctly in filter
mount.
Insert filter correctly in filter
mount.