Installation of the AxioCam MRc5
32
B 48-0028 e 10.2003
Resetting the AxioCam MRc5 firmware
If the color of the LED on the back of the AxioCam MRc5 does not
immediately change from red to green but goes out after the insertion of the
FireWire cable, this means that the AxioCam MRc5’s firmware is not
functioning properly.
This error can occur, for example, if the data transmission or power supply to
the camera has been cut off during the firmware recording procedure.
If this happens, you need to record the firmware of the AxioCam MRc5 again:
Connect the AxioCam MRc5 to your computer in accordance with the
description in the section "Setting up the camera system". The LED on
the back of the camera lights up in red briefly and then goes out.
Press the button next to the expansion socket of the AxioCam MRc5
using a pointed object (e.g. a pencil). Keeping the button depressed,
briefly remove and then reinsert the FireWire connector on the camera.
Now the LED lights up red and then changes to green.
Do not press down hard on the button. This can damage the
camera’s electronics.
Update the firmware of the AxioCam MRc5 by running the installation
program. Here follow the description in the section "Installing the camera
driver using the setup program". At the end of the installation program
the firmware of the AxioCam MRc5 must be updated. Follow the
instructions on the monitor.