Maintenance
Section
6
2.
Use the [UP], [DOWN] and [DIGIT] soft keys to enter the platinization time required in
minutes. Six minutes is usually sufficient for most electrodes.
3.
Immerse the cell in the platinizing solution (YSI 3140). Make sure that both electrodes
are submerged.
4.
Press [ENTER] to start the platinization process. The following screen, showing the time
remaining and the polarity of the polarization current, will be displayed.
PLATINIZE
5.8 MIN remaining time
+ platinization polarity
PRESS MODE TO ABORT PLATINIZATION
5.
After the platinization process is complete, remove the cell from the platinizing solution.
If you want to stop the platinization before the timer reaches zero, press the [MODE] key
to abort.
6.
Thoroughly rinse the cell with distilled or deionized water.
7.
Promptly return the platinizing solution to its container.
YSI, Incorporated
Model 3200
38
Содержание 3200
Страница 2: ......
Страница 30: ...System Configuration and Operation Section 3 YSI Incorporated Model 3200 26 ...
Страница 40: ...Diagnostics Section 5 YSI Incorporated Model 3200 36 ...
Страница 44: ...Troubleshooting Section 7 YSI Incorporated Model 3200 40 ...
Страница 48: ...Principles of Operation Section 8 3252 3253 3254 3255 3256 YSI Incorporated Model 3200 44 ...
Страница 56: ...Principles of Operation Section 8 YSI Incorporated Model 3200 52 ...
Страница 59: ...Warranty and Repair Section 9 YSI Incorporated Model 3200 55 ...
Страница 70: ...Accessories and Replacement Parts Section 10 YSI Incorporated Model 3200 66 ...
Страница 74: ...Required Notice Appendix B YSI Incorporated Model 3200 70 ...
Страница 76: ...RS232 Connections Appendix C YSI Incorporated Model 3200 72 ...
Страница 82: ...Conversion Factors Appendix E YSI Incorporated Model 3200 78 ...
Страница 85: ...Glossary of Terms Appendix F YSI Incorporated Model 3200 81 ...