System Configuration and Operation
Section 3
[0] # Configuration number
CONFIG
[CONFIG 0] Title
0.0000uS offset
[Single]
Cell
Cal
[1.00]
K
[Linear] Comp
[1.91] %/°C Comp val
[0.68] T.D.S. mult
[25.0] °C Comp Temp
ENTER TO SEND CONFIGURATION
UP
DOWN
DIGIT
NEXT
Press the [UP] or [DOWN] soft key to select one of the six (0-5) stored configurations. Each
configuration contains the single cell constant, zero offset, multiple cell constants, total dissolved
solids (TDS) multiplier,
linear temperature compensation coefficient
and
non-linear temperature
compensation data.
To change the currently selected configuration, press [NEXT] to move the cursor to the field you
want to change. Then, use the [UP], [DOWN] and [DIGIT] soft keys to change the field, except
Offset. The Offset field can only be changed from the Offset menu. The following table shows
the choices for each field.
Example: To select the standard non-linear temperature compensation curve for ultrapure water,
press [NEXT] three times, to move the cursor to the Comp field, then, press the [UP]
or [DOWN] soft key to cycle through the selections until Pure is displayed. Press
[MODE] two times to exit to the Operation menu.
Field
Range / options
Description
Title
A - Z or 0 - 9
User defined name for this configuration.
Cell Cal
Single or Multi
Selects between a single fixed cell constant
and multiple cell constants.
Comp
Linear, N. Linear, Pure, Natural
or Off
Selects between a linear temperature
compensation coefficient, non-linear
temperature compensation curve, standard
non-linear temperature compensation curve
(pure or natural) or no temperature
compensation.
TDS mult
0.55 - 0.90
Factor used to calculate the total dissolved
solids from the conductivity.
uS offset
0.0000 - 0.0150
Zero offset. Used only when measuring
conductance below 10uS to offset cell lead
insulation resistance. This value can only be
changed from the Offset menu.
K
0.001 - 100.999
Single fixed cell constant value (K). This
value is only used when the Cell Cal field is
set to Single.
YSI Incorporated
Model 3200
13
Содержание 3200
Страница 2: ......
Страница 30: ...System Configuration and Operation Section 3 YSI Incorporated Model 3200 26 ...
Страница 40: ...Diagnostics Section 5 YSI Incorporated Model 3200 36 ...
Страница 44: ...Troubleshooting Section 7 YSI Incorporated Model 3200 40 ...
Страница 48: ...Principles of Operation Section 8 3252 3253 3254 3255 3256 YSI Incorporated Model 3200 44 ...
Страница 56: ...Principles of Operation Section 8 YSI Incorporated Model 3200 52 ...
Страница 59: ...Warranty and Repair Section 9 YSI Incorporated Model 3200 55 ...
Страница 70: ...Accessories and Replacement Parts Section 10 YSI Incorporated Model 3200 66 ...
Страница 74: ...Required Notice Appendix B YSI Incorporated Model 3200 70 ...
Страница 76: ...RS232 Connections Appendix C YSI Incorporated Model 3200 72 ...
Страница 82: ...Conversion Factors Appendix E YSI Incorporated Model 3200 78 ...
Страница 85: ...Glossary of Terms Appendix F YSI Incorporated Model 3200 81 ...