Getting Started
Section 2
3. Connect the Cell
Plug the cell (YSI 3200 series) into the connector on the back of the instrument marked cell.
NOTE: If you are using a YSI 3400 series cell, you will need the following:
•
3232 adapter
•
YSI 3220 or 700 series temperature probe
4. Configure the 3200
Next, you
must
configure the 3200 with the correct cell constant of the cell you are using.
From Operation mode, press the [CELL] soft key to enter the Cell menu.
[0] # Configuration number
CELL
[CONFIG 0] Title
0.0000uS offset
[Single] Cell Cal
[1.00] K
[Linear] Comp
[1.91] %/°C Comp val
[0.68] T.D.S. mult
[25.0] °C Comp Temp
ENTER TO SEND CONFIGURATION
CONFIGURE
CAL K
TEMP
PLATIN
Then, press the [CONFIGURE] soft key to enter the Configure menu. Configure is used to select
and configure the current cell and temperature compensation. The 3200 stores six configurations
numbered 0 to 5.
K field (cell constant)
[0] # Configuration number
CONFIG
[CONFIG 0] Title
0.0000uS offset
[Single]
Cell
Cal
[1.00]
K
[Linear] Comp
[1.91] %/°C Comp val
[0.68] T.D.S. mult
[25.0] °C Comp Temp
ENTER TO SEND CONFIGURATION
UP
DOWN
DIGIT
NEXT
Press the [UP] or [DOWN] soft key until you have selected configuration 0. The default
configuration is a single fixed cell constant of K = 1, linear temperature compensation corrected
to 25°C using a coefficient of 1.91%/°C.
Press the [NEXT] soft key to move the cursor (underline) to the Title field. Using the [UP],
[DOWN] and [DIGIT] soft keys, you may enter a name for this configuration.
Press the [NEXT] soft key four times to move the cursor to the K field. Using the [UP], [DOWN]
and [DIGIT] soft keys, enter the manufacturer’s stated cell constant for the cell that you are using
(K= 0.1, 1.0, 5.0, 10.0 etc.).
Press the [MODE] key two times to return to the Operation menu.
YSI Incorporated
Model 3200
9
Содержание 3200
Страница 2: ......
Страница 30: ...System Configuration and Operation Section 3 YSI Incorporated Model 3200 26 ...
Страница 40: ...Diagnostics Section 5 YSI Incorporated Model 3200 36 ...
Страница 44: ...Troubleshooting Section 7 YSI Incorporated Model 3200 40 ...
Страница 48: ...Principles of Operation Section 8 3252 3253 3254 3255 3256 YSI Incorporated Model 3200 44 ...
Страница 56: ...Principles of Operation Section 8 YSI Incorporated Model 3200 52 ...
Страница 59: ...Warranty and Repair Section 9 YSI Incorporated Model 3200 55 ...
Страница 70: ...Accessories and Replacement Parts Section 10 YSI Incorporated Model 3200 66 ...
Страница 74: ...Required Notice Appendix B YSI Incorporated Model 3200 70 ...
Страница 76: ...RS232 Connections Appendix C YSI Incorporated Model 3200 72 ...
Страница 82: ...Conversion Factors Appendix E YSI Incorporated Model 3200 78 ...
Страница 85: ...Glossary of Terms Appendix F YSI Incorporated Model 3200 81 ...