5-11
Enter WHITE end point in seconds
> 30
Press [ENTER] to confirm and return to Measurement parameter setup.
Measurement parameter setup
1-SampleSize 2-CalMethod 3-Black 4-White
You have now completed MeasParameter setup. Press [0] to return to the Select setup level.
Select setup: 1-General 2-MeasParameter
3-RunMode 4-Report 5-PrntSetup 6-Default
Press [3] for RunMode. The display will read:
RUN mode setup: 1-SampleProtocol
2-AutoStandby 3-AutoCal 4-Monitor
In this menu, you select parameters related to sampling protocol, set parameters that control
automatic switching to standby mode and select parameters that will trigger autocalibrations. In
Sampling Protocol, you can select the sample station (position), activate replicate sampling and
sample identification systems, and set the position to which the Sipper descends when sampling
at the Manual Station.
Press [1] for SampleProtocol. The display will read:
Sampling protocol setup: 1-SipperHeight
2-Replicates 3-ID 4-Station# 5-Turntable
Press [1] for SipperHeight. This specifies to what vertical position the Sipper descends when
sampling at the Manual Station (#3). For example, with long test tubes and the fluid level near
the bottom of the test tube, use "Low" setting.
Select manual sipper height: Medium
1-Low 2-Medium 3-High
Press [ENTER] to confirm and return to Sampling protocol setup.
Sampling protocol setup: 1-SipperHeight
2-Replicates 3-ID 4-Station# 5-Turntable
Press [2] for Replicates. The default setting is No. The display prompts reflect messages for
single sampling and single calibrations. If Yes is confirmed, you will be prompted at the display
each time you press [SAMPLE] or [CALIBRATE] to enter how many times the sample or cal
solution should be run for that one command. The number that you assign to Replicates must be
between 1 and 999 (samples or calibrations). During RUN Mode, you may cancel the assigned
number of replicate cycles by pressing [CANCEL].
As an example, Replicates allows you to configure your system to run a sample in the Test Tube
Holder unattended in duplicate or triplicate. Replicates also allows you to check calibration
stability by assigning some number of consecutive calibrations to be run without requiring
repeated key commands of [CALIBRATE].
Содержание 2700 SELECT
Страница 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Страница 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Страница 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Страница 49: ...3 10 ...
Страница 95: ...5 26 ...
Страница 103: ...6 8 ...
Страница 117: ...7 14 ...
Страница 143: ...9 18 ...
Страница 153: ...D 2 ...
Страница 155: ...E 2 ...
Страница 159: ...F 4 ...
Страница 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Страница 161: ...G 2 ...
Страница 163: ...H 2 ...