4-11
)
This is a direct reading of choline in solution at the enzyme sensor. The enzyme Choline
Oxidase is immobilized in the YSI Choline Membrane.
C 2O
2
Chol Oxidase
2H
2
O
2
+ Betaine
System Buffer
YSI 2357
Calibrator Std
YSI 2772
Linearity Std
YSI 2773
Black Probe
YSI 2771(Choline Oxidase Membrane)
White Probe
n/a *
Sample Size
25
µ
L
Cal Station
Station #1 Cal Well
Black Probe Parameters
Chemistry
Choline
Unit of Conc
mg/L
Cal value
175
End Point
30 sec
* If you have a dual channel unit, either Black or White, or both Black and White probes can be
configured for this chemistry.
Note: See Appendix A if concentration unit conversion is required.
Special Considerations:
»
If sample dilution is required, use reagent water or 2357 Buffer.
»
Although the YSI 2772 Choline calibrator solution is prepared using choline bitartrate,
the concentration is expressed as mg/L of choline cation. If you prefer to express the
sample as a salt of choline, you must enter an “adjusted” calibration value when
configuring your instrument measurement parameters.
For example, to express results as a choline hydroxide value, program the calibrator
value as 204 mg/L when using the YSI 2772 Choline Standard calibrator solution. Your
results will now reflect the concentration of choline hydroxide in the sample.
Assumptions and calculations are as follows:
choline hydroxide FW = 121.2
free choline FW = 104.2
(121.1
÷
104.2) x 175 mg/L = 203.55 = 204 mg/L (rounded up)
Equivalent values for other choline salts may be calculated in a similar manner.
Potential Interferences. Tests with equimolar concentrations of various vitamins, and other
ingredients of nutritionals, showed little or no interference. On a scale where Choline = 100%,
those substances that produced more than 0.5% response were Riboflavin (1.1%), Pyridoxine
(1.5%), Ascorbic Acid (0.8%), and Thiamine (1.3%). Vanillin, an artificial flavor ingredient,
may produce interference levels exceeding 10% (mole/mole), and should be tested separately
for your individual formulations if known to be present.
Содержание 2700 SELECT
Страница 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Страница 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Страница 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Страница 49: ...3 10 ...
Страница 95: ...5 26 ...
Страница 103: ...6 8 ...
Страница 117: ...7 14 ...
Страница 143: ...9 18 ...
Страница 153: ...D 2 ...
Страница 155: ...E 2 ...
Страница 159: ...F 4 ...
Страница 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Страница 161: ...G 2 ...
Страница 163: ...H 2 ...