1-6
Lactose
YSI Membrane:
2702
Detection Range:
0-25 g/L
Calibration Point:
5.00 g/L
Linearity Check Point:
25.0 g/L
Precision (CV,n=10):
2%
Linearity (0 to Cal Point):
±
2%
Linearity (Cal to Range Max):
±
5%
Typical Membrane Working Life:
10 days
Membrane O-ring Color:
White
Sucrose
YSI Membrane:
2703
Detection Range:
0-25 g/L
Calibration Point:
5.00 g/L
Linearity Check Point:
25.0 g/L
Precision (CV,n=10):
2%
Linearity (0 to Cal Point):
±
2%
Linearity (Cal to Range Max):
±
5%
Typical Membrane Working Life:
10 days
Membrane O-ring color:
Blue
Choline
YSI Membrane:
2771
Detection Range:
0-450 mg/L
Calibration Point:
175 mg/L
Linearity Check Point:
450 mg/L
Precision (CV,n=10):
2%
Linearity (0 to Cal Point):
±
2%
Linearity (Cal to Range Max):
±
5%
Typical Membrane Working Life:
7 days
Membrane O-ring color:
Orange
L-Glutamine*
YSI Membrane:
2735
Detection Range:
0-8 mmol/L
Calibration Point:
5.00 mmol/L
Linearity Check Point:
8.00 mmol/L
Precision (CV,n=5 ):
4%
Linearity (0 to Cal Point):
±
4%
Linearity (Cal to Range Max):
±
5%
Typical Membrane Working Life:
5 days
Membrane O-ring color:
Magenta
*Specs shown are for simultaneous measurement of glutamine and glutamate.
Содержание 2700 SELECT
Страница 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Страница 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Страница 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Страница 49: ...3 10 ...
Страница 95: ...5 26 ...
Страница 103: ...6 8 ...
Страница 117: ...7 14 ...
Страница 143: ...9 18 ...
Страница 153: ...D 2 ...
Страница 155: ...E 2 ...
Страница 159: ...F 4 ...
Страница 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Страница 161: ...G 2 ...
Страница 163: ...H 2 ...