2-10
Press [1] for Lower sipper for fine alignment. The Sipper will move closer to the small opening
in the stainless steel cone.
Fine align sipper then select
0-Exit 1-Test sipper position
The tip of the Sipper should be exactly centered above this opening. If necessary, loosen the
adjustment screw again and position the Sipper exactly. Retighten the adjustment screw.
Figure 2.7
Sipper Adjustment Position
Press [1] for Test Sipper position. The Sipper will descend into the Sample Chamber. The
Sipper should not contact the stainless steel cone. If the Sipper position is still not exactly right,
press [1] and readjust it once more, as described above. IT IS VERY IMPORTANT THAT
THE SIPPER BE ACCURATELY ADJUSTED.
Select 1 to restart check sipper cycle
0-Exit 1-Home sipper position
After adjustment is complete, press [0] to exit and return to the Select Service menu level.
Select service: 1-Sipper 2-Buffer 3-Cal
4-Stir speed 5-Monitor 6-Turntable