5-7
Select number of data bits: Seven bit
1- Seven bit 2-Eight bit
Select parity: Odd
1-None 2-Even 3-Odd 4-Low 5-High
Select number of stop bits: One bit
1-One bit 2-Two bits
Select handshaking type: RTS/CTS
1-RTS/CTS 2-XON/XOFF 3-None
If you choose XON/XOFF, you will go one menu level deeper to enter or confirm XON and
XOFF values.
Enter XON character
> 17
Press [ENTER] to confirm.
Enter XOFF character
> 19
Press [ENTER] to confirm and return to RS-232 setup.
RS-232 setup 1-Baud 2-Data 3-Parity
4-Stop 5-Handshake 6-Configuration
If you press [6] for Configuration, you will see the following.
Select configuration: Non-multidrop
1-Non-multidrop 2-Multidrop
The configuration option allows operation in either a point-to-point environment or in a
multidrop environment with an address (ID) assigned to your 2700 SELECT unit(s). The
default configuration is Non-multidrop. If you press [2] for Multidrop, you will see the
following.
Enter node address
> 38
Either enter another address # or accept the default, then confirm with [ENTER]. The node
address is only used when networking 2700s in a multi-instrument system. The 2700 responds
only to commands prefixed with the ID matching the address assigned from this menu.
Содержание 2700 SELECT
Страница 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Страница 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Страница 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Страница 49: ...3 10 ...
Страница 95: ...5 26 ...
Страница 103: ...6 8 ...
Страница 117: ...7 14 ...
Страница 143: ...9 18 ...
Страница 153: ...D 2 ...
Страница 155: ...E 2 ...
Страница 159: ...F 4 ...
Страница 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Страница 161: ...G 2 ...
Страница 163: ...H 2 ...