Additional remote control commands related to changing the external monitoring parameters
and aborting a turntable run are described below.
Monitor Parameter Modification Commands (MP, MT, MR, MO, MS). You may use the
commands described below to change monitor setup parameters remotely.
MPt
change external pump purge time to t, where t is in seconds
MTt
change monitor time interval to t, where t is in minutes
MRt
change precal time interval to t, where t is in minutes
MOt
change postcal time interval to t, where t is in minutes
MSs
change monitor station number to s
Abort Turntable Sample Command (PA). You may use this command to abort turntable
samples in progress. Recall, you initiate the turntable sampling using the PS command. You
cannot use the 2700 SELECT [CANCEL] key to abort the run, since the 2700 SELECT keypad
is disabled in remote control mode.