4-15
#
This is a direct reading of glutamine in solution at the enzyme sensor. Two enzymes are co-
immobilized in the YSI Glutamine Membrane: Glutaminase and Glutamate Oxidase.
L-Glutamine
Glutaminase
L-Glu NH
3
L-Glu O
2
Glut Oxidase
H
2
O
2
+
α
-Ketoglu NH
3
Through this chain of reactions the amount of hydrogen peroxide liberated is directly
proportional to the amount of glutamine.
System Buffer
YSI 2357
Calibrator Std
YSI 2736 (5.00 mmol/L)
Linearity Std.
YSI 2737 (8.00 mmol/L)
Black Probe
YSI 2735 Glutamine Membrane
White Probe
n/a *
Sample Size
20
µ
L
Cal Station
Station #2
Black Probe Parameters
Chemistry
Glutamine
Unit of Conc
mmol/L
Cal value
5.00
End Point
30 sec
* If you have a dual channel unit, either Black or White, or both Black and White probes can be
configured for this chemistry.
Note: See Appendix A if concentration unit conversion is required.
Special Considerations:
»
If sample dilution is required, use reagent water or 2357 buffer.
The sample must be glutamate-free, or at least contain levels low enough not to interfere with
the glutamine reading. Since the glutamine membrane contains glutamate oxidase, glutamate
will produce a probe signal. See Section 4.17 to measure glutamate and glutamine
simultaneously.
Содержание 2700 SELECT
Страница 1: ... Biochemistry Analyzer Who s Minding the Planet find out at ...
Страница 4: ...iii A 1 B 1 C 1 D 1 0 1 E 1 2 F 1 3 4 5 2 G 1 6 7 H 1 I 1 ...
Страница 43: ...3 4 Figure 3 2 Sampling Stations Figure 3 3 The Test Tube Holder Pivoted Out ...
Страница 49: ...3 10 ...
Страница 95: ...5 26 ...
Страница 103: ...6 8 ...
Страница 117: ...7 14 ...
Страница 143: ...9 18 ...
Страница 153: ...D 2 ...
Страница 155: ...E 2 ...
Страница 159: ...F 4 ...
Страница 160: ...G 1 Figure G 1 Line Power Cord and Plug Wiring ...
Страница 161: ...G 2 ...
Страница 163: ...H 2 ...