Monitor the entire priming procedure continuously until the tubing of the infusion set has been filled without any
air bubbles and until insulin emerges from the end of the tubing. If there are still air bubbles in the tubing or the adapter,
repeat the priming procedure. Air bubbles can lead to incorrect dosages of insulin, which can cause hypoglycemia
or hyperglycemia (read warning in the box on page 7).
The “Prime tubing” and “Prime cannula” functions must never be used for bolus delivery – their sole purpose is to prime
the tubing and cannula without bubbles. Misuse of these functions can lead to incorrect dosages of insulin, which can
cause hypoglycemia or hyperglycemia (read warning in the box on page 7).