Priming not completed
The “Priming not completed” warning appears if:
the drive is fully extended (cartridge empty) without a plunger having been detected.
the priming procedure is canceled by the user during plunger detection.
the priming procedure is canceled by an alarm.
the force on the plunger holder decreases during the priming procedure, for example because the cartridge
has been removed or a wrong cartridge has been inserted.
Confirm by tapping and perform the priming function again. Make sure the cartridge has been inserted properly and
that the adapter is connected to the insulin pump properly. Follow the procedure on page 99.