146
8.1 Incidents
Air bubbles in the cartridge or
in the infusion set
Check your blood glucose level. Disconnect the infusion set from your body and
prime the infusion set without any air bubbles in accordance with the procedure
on page 99.
“Basal rate profile A” icon is not
available (shown in gray)
If the temporary basal rate function is activated, the Basal rate profile A function
is not available and it is not possible to change the basal rate profile.
“Basal rate profile B” icon is not
available (shown in gray)
If the temporary basal rate function is activated, the Basal rate profile B function
is not available and it is not possible to change the basal rate profile.
Blind bolus function is not
available
Make sure the blind bolus function is activated and the insulin pump is in run mode.
The blind bolus function is not available in stop mode. The blind bolus function
cannot be performed if the cartridge is empty, nor if the remaining insulin amount in
the cartridge or the bolus limit setting is smaller than the bolus increment setting.
The blind bolus function cannot be performed either if a threaded rod return has been
performed without any subsequent priming or if the priming function has not been
completed properly.
Содержание YpsoPump
Страница 1: ...More freedom More confidence With Ypsomed User Guide Insulin Pump ...
Страница 2: ......
Страница 26: ...26 3 3 The status screen appears Basal rate delivery starts according to the programmed profile ...
Страница 54: ...54 4 4 The status screen appears with the basal rate originally programmed ...
Страница 129: ...129 3 3 Confirm by tapping The YpsoPump vibrates briefly and the orientation of the screen is rotated by 180 ...
Страница 202: ...Notes 202 ...
Страница 203: ...203 ...