Confirm that you have discon-
nected the infusion set from your
body by removing the tubing cap from
the cannula base and confirm by
tapping . The YpsoPump
®
vibrates
briefly, the threaded rod advances
up to the plunger of the cartridge and
the infusion set is primed with the set
amount of insulin.
5/6:
During the priming procedure,
keep the insulin pump in an upright
position, with the adapter facing up-
wards and tap the insulin pump lightly
on your open palm to remove the air
bubbles from the infusion set. The
screen counts the delivered priming
volume upwards until the set value
has been reached. You can cancel
the priming procedure at any time by
tapping .
If the temporary basal rate function or a bolus is activated, the prime infusion set function is not available.
The prime infusion set function also cannot be executed if the cartridge is empty.