![Ypsomed YpsoPump Скачать руководство пользователя страница 161](http://html.mh-extra.com/html/ypsomed/ypsopump/ypsopump_user-manual_915220161.webp)
161
Alarms given are primarily tactile and audible and there are 4 escalation levels. The levels escalate at intervals
of 5 minutes. If an alarm is confirmed but not dealt with, it is given again by audible and tactile means after 30 minutes.
An electronic error is the exception. That is indicated directly and permanently by audible and tactile means.
Alarms that do not involve an electronic error appear on the status screen and have to be confirmed. To do so, the screen
has to be switched on first using the function button. Alarms always lead to cancelation of insulin delivery. The alarm
remains on the status screen until the actions to deal with the alarm have been performed on the pump. An electronic error
is the exception. It is dealt with in accordance with the procedure on page 171.
Escalation levels alarms:
Level 1: Tactile
Level 2: Tactile and low audible signal
Level 3: Tactile and high audible signal
Level 4: Tactile and higher audible signal