The “No insulin” alarm appears if after a threaded rod return no priming procedure is performed within 5 minutes
or if the priming procedure failed. The “No insulin” alarm only appears if the insulin pump is in run mode.
Confirm by tapping . Insert a cartridge and perform a priming procedure. Follow the procedure on page 95.