2/7:
You are now in the blind bolus menu. Here you can
set the bolus amount required. Begin with entering an
amount within 4 seconds. The elapsed time is displayed
to you by dots turning gray.
If you fail to set a bolus amount within 4 seconds, the
blind bolus menu disappears automatically and
programming is canceled. This is indicated by three short
vibrations.
3/7:
Set the bolus amount required by giving the function
button a short press for each step of increase. The insulin
pump confirms each function button press with a short
vibration. For each function button press, the bolus
amount is increased depending on the bolus increment
you have set. The maximum bolus amount is 30.0 U.
Function button
75
Giving the function button a long press may launch the blind bolus menu accidentally, if the blind bolus
function was previously activated in the settings. If you do not react within 4 seconds, the blind bolus menu
disappears. It is not possible to program the blind bolus in the main menu or in a submenu.
Содержание YpsoPump
Страница 1: ...More freedom More confidence With Ypsomed User Guide Insulin Pump ...
Страница 2: ......
Страница 26: ...26 3 3 The status screen appears Basal rate delivery starts according to the programmed profile ...
Страница 54: ...54 4 4 The status screen appears with the basal rate originally programmed ...
Страница 129: ...129 3 3 Confirm by tapping The YpsoPump vibrates briefly and the orientation of the screen is rotated by 180 ...
Страница 202: ...Notes 202 ...
Страница 203: ...203 ...