7.6 Preparing for storage
If you will not be using the YpsoPump
®
for a longer period, set your insulin
pump into stop mode and remove the
battery from the insulin pump to avoid
any damage caused by a leaking
battery. That puts the insulin pump in
storage status. In the storage status
the settings remain intact (basal rate,
bolus increment, etc.).
When preparing your YpsoPump
®
for storage, remove the cartridge and the infusion set and dispose of the
tubing and cannula base in a safety bin.
After a restart all you have to do is
perform the self-test and reset the
time of day and date. After that,
the insulin pump is in stop mode. A
threaded rod return followed by
priming has to be performed to put
the insulin pump back into operation.
Follow the procedure on page 31.