www.youfibercable.com
Email:[email protected] MSN:[email protected]
30
17) 1>The heat LED
doesn’t work, but it can
heat; or the index light
,
buzzer and heater don’t
work
2> The heat LED and
buzzer work well, but the
heater doesn’t work.
1.
The heat key
is damaged.
2.
The heat key
is stuck by
something
hard.
2.
Press “Heat” several
times, If it doesn’t
work, contact us.
All problems are
caused by inner
circuit fault
probably.
If you couldn’t
solve the
problems,
please sent it
to us with
detailed
explanation.
18).Can’t achieve arc in
fusion and current test or
something wrong happen.
1. When press the
“RUN” and the
machine is in mode of
auto splice,” HEAT”
is pressed by mistake.
2. When press
“TEST” and the
machine is in arc test,
“HEAT” is pressed by
mistake.
1. Don’t press “HEAT”
when pressing the
“RUN” and the machine
is in mode of auto
splice.
2. Don’t press “HEAT”
when pressing “TEST”
and the machine is in
arc test.
15
Addenda E Introduction of power supply modules
KL-01-26L7 PS modules has inner Lithium battery, 11.1V 6.6Ah. It is fixed on the bottom of
machine with four screws and easy to replace with special design.
1)
Check electric quantity of battery and charging mode explanation:
A. Press “PUSH” to show electric quantity when machine is off. More green lights on the
left side show more electric quantity. The red light stands for charging. Please refer to
following table:
Battery power
≥80%
Battery voltage
≥12.2V
Battery power
≥60%
Battery voltage
≥11.8V
Battery power
≥40%
Battery voltage
≥11.4V
Battery power
≥15%
Battery voltage
≥11.0V
Battery power
≥10%
Battery voltage
≥11.0V
Electric
quantity100%
Electric quantity
90%
Electric quantity
60%
Electric quantity
30%
Electric quantity
<
10%