5) Press “RUN”,
Fiber image of one
Side moved vertically,
and could not splice.
1.There is dust in the
V-groove, which
leads to the position
of fiber is too higher
than another side.
2. There is dust on
lens, light or reflector.
1.Wipe V-groove at
single direction with tool
several times.
2.Clean lens, prism
and two lights with
cleanly cotton several
times.
3. If it doesn’t work,
contact us.
6)After splicing,
displayed big loss or
failed splicing.
1.
The fiber is dirty,
the end-face is
unqualified, or there is
fault in the cleaver.
2. There is dust on the
lens, light or reflector.
1.Adjust fiber cleaver,
make qualified fiber
end-face.
2.
Clean the lens,
reflector and two lights
with cleanly cotton
several times.
3. If it doesn’t work,
contact us.
All problems are
caused by inner
circuit fault
probably.
If you couldn’t
solve the
problems,
please sent it
to us with detailed
explanation.
7)Always display
bad fiber end-face of one
side.
1.
The value of
“Setting end-face”
is too small.
2. There is dust on the
Lens, lights or
reflector..
3.The light is dim.
4.There is dust in V-
groove or fiber isn’t
put properly.
Enter the menu; add the
value of “Setting end-
face”. Clean the lens,
the light and the
reflector, clean the V-
groove and try again.
. If it doesn’t work,
contact us.
8)The value of tested
current is too big or too
small.
1.The value of “fusion
current” are too large
or too small.
2.
Too much
sediment on the
electrode.
3.
The relative
position between
fiber and arc has
changed.
4.
The storage
battery doesn’t
work well.
5.
The environment
has changed.
1.
Enter the menu and
clean electrode
several times.
2.
Or select “electrodes
aging” and check
whether the arc is
steady.
Then select Program 3
to test arc again.
3.
Test the current
until it is proper.
4.
If it doesn’t work,
contact us.