5)
If “Not qualified end-surface” or “Reload fiber” is displayed, KL-280 will stop
working. Please remake fiber’s end-surface or refer table 2. (Consult 6.2.4 and 6.2.5)
☆
Notice: 1> Please do not press “HEAT” after you press “RUN” and when machine in status of
auto splice, otherwise will have error.
2>Fusion splicer will estimate loss one time only after splicing automatically.
3> Only the electrode is well, arc parameter is optimized, fiber character is
identical, then fusion loss is of reference.
4> Best to arc test before splicing.
6.3.5 Splicing partial core fiber
If loss is unsteady during splicing, the measures are as follows:
1>
Check whether the arc is steady and whether the parameter is proper.
2>
If opposite position is steady and the parameter is proper, maybe fiber’s core
is partial. The solution is minish the value of “fusion time” and “fusion push”.
3>
If method<2> is useless, to obtain best loss, you must mark direction angle on the
fiber to test fiber’s fusion.
6.3.6 Evaluating fusion quality
Fusion quality is judged by estimating loss, fusion shape and fusion process.
Only combine them together, and then the evaluation is external.
1>
If fusion process is steady and the spliced fiber’s shape as follows as Table 3,
then the fusion status is well.
Table 3
Fiber’s cord is accord with shape
Spot or abrasion, be care of cleaning and
cleaving end-surface
white
line
Because of optics, there’s no effect to joint
Different eccentricity between two fibers
Different external diameter between two fibers
2>If the spliced fiber’s shapes are as follows as Table 4, you must splice over again even
the loss value is very low.