IM 701210-07E
38
MODE (ACTION)
Auto mode
Section 6.1
If the trigger condition is met within the
timeout period (approximately 50 ms), the
waveform is updated on each trigger
occurrence. If the trigger condition is not met
after the timeout period elapses, the waveform
is automatically updated.
Auto level mode
Section 6.1
Within the timeout period (approximately 1 s),
the waveforms are displayed in the same
fashion as in the auto mode. If a trigger is not
activated within the timeout period, then the
center value of the amplitude of the trigger
source is detected, and the trigger level is
changed to that value. A trigger is activated
using the new value, and the displayed
waveforms are updated.
Normal mode
Section 6.1
The waveform display is updated only when
the trigger conditions are met.
Single mode
Section 6.1
The display is updated once when the trigger
conditions are met and the waveform
acquisition stops.
Single (N) mode
Section 7.5
When the trigger conditions are met,
waveform acquisition is repeated the specified
number of times, and the waveform acquisition
stops. After waveform acquisition is finished,
all the waveforms are displayed.
Log mode
Section 6.1
In this mode, the trigger settings are disabled.
The specified record length of data is acquired
once when acquisition is started, and the
displayed waveforms are updated.
Single (N) count
Section 7.5
Set the number of waveform acquisitions for
single (N) mode.
Auto Level Mode and Single (N) Mode
• Auto Level Mode
• Single (N) Mode
Trigger Mode during DL750P Recorder Mode
If Chart Recorder mode (see page 56) is selected on the
DL750P, select the trigger mode from Auto, Log,
Single, and Repeat. If X-Y Recorder mode is selected,
the trigger mode cannot be set.
Note
1/2 the amplitude
1/2 the amplitude
Trigger level Amplitude
1st
acquisition
2nd
acquisition
Nth
acquisition
Setup Menu Items