background image

IM 701210-07E

14

Main Functions of the DL750/DL750P

Realtime Recording to the Internal Hard Disk (/C8 Option)

Section 7.7, “Realtime Recording to the Internal Hard Disk (Optional)” in the User’s Manual Part 1

Measured data can be saved in realtime over long periods to the internal hard disk (requires the /C8 option).  On the internal hard disk,
half the total size of the disk is allocated beforehand to be used as area for realtime recording.  You can set the size of the realtime
recording area in the range of 30% to 70% of the entire capacity of the internal hard disk.  The saved data can be loaded, and the
measured data can be managed and analyzed on a PC.

Input signal

Copy

Acquisition

memory

Measurements other
than realtime recording

(normal measurement)

Realtime recording

Load waveform data

Save and Load setup data

PC

Realtime

recording area

(30% to 70% of the 

entire capacity)

Data

storage area

Copy

SCSI
GP-IB interface
Ethernet interface
USB interface

Internal hard disk

Recording in Recorder Mode (Realtime Recording) (DL750P Only)

Chapter 9, “Recording in Recorder Mode (Realtime Recording)” in the User’s Manual Part 1

The DL750P is equipped with a recorder mode in which waveforms and numeric values can be recorded on a built-in A4-size printer. The
recorder mode is set using the RECORDER key on the DL750P front panel. There are two recorder modes.

Chart Recorder Mode

When waveform acquisition is started, waveforms and numeric values can be recorded in realtime on the built-in printer. Because the
data is saved to the internal memory at the same time, the required sections can be redisplayed such as by using the search & zoom
function after the recording is completed. The stored data can be handled in the same fashion as normal data such as saving the data to a
file, performing cursor measurements, automated measurement of waveform parameters, normal statistical processing, and search & zoom.

• T-Y Waveform Recording

T-Y waveforms are recorded at a specified chart speed. If shot recording is specified, the recording stops automatically when the specified
length of data is recorded. The data can be reprinted by changing the print conditions, and a PDF file of the reprinted image can be created.

• Numeric Value Recording

The numeric values are recorded at specified time intervals. Up to 16 channels can be recorded at once.

X-Y Recorder Mode (X-Y Waveform Recording)

When waveform acquisition is started and stopped, the X-Y waveforms during the waveform acquisition period can be recorded. If auto
print is specified, the recording of the X-Y waveforms starts when waveform acquisition is stopped.  A PDF file of the reprinted image
can be created.

Chart speed or 
recording interval

Recording graticule

Acquisition mode

Trigger mode

Number of data points that can 
be saved to the internal memory

Functions that cannot
be used simultaneously

Chart Recorder Mode

T-Y waveform recording

Chart speed: 

10 mm/h to 20 mm/s
Horizontal axis (time axis): 10 mm/div

Vertical axis: 10 mm/div or [1 division of 

the 10 equally divided recording area]/div

Envelope (the mode cannot be changed with the ACQ key, 

envelope and normal can be changed with the MISC key)

Auto, Log, Single

1

, Repeat

1

 (1: Only during T-Y waveform recording)

Up to 1000 div (varies depending on the chart speed) 

    

History memory, GO/NO-GO determination, dual capture, waveform computation, 

and search & zoom (only during X-Y Recorder mode)

Numeric value recording

Recording interval: 

1 s to 60 min

—   

X-Y Recorder mode 
(X-Y waveform recording)

X-axis/Y-axis: 10 mm/div or 

[1 division of the 10 equally 

divided recording area]/div

Normal 

(the mode cannot be 

changed with the ACQ key)

Fixed to 1 MW

Chart Recorder mode and X-Y Recorder mode

Содержание ScopeCorder DL750P

Страница 1: ...ScopeCorder IM 701210 07E 3rd Edition...

Страница 2: ...those that appear in the actual screen Every effort has been made in the preparation of this manual to ensure the accuracy of its contents However should you have any questions or find any errors ple...

Страница 3: ...the Screen 18 Changing the Waveform Display Conditions 19 Changing the Trigger Setting 22 Measuring Waveforms 24 Zooming the Waveform along the Time Axis 25 Printing Saving Waveforms 26 Setup Menu Ite...

Страница 4: ...is Time Axis Vertical Axis Trigger Waveform Acquisition Computing Analyzing and Searching Waveforms Computation Cursor Measurement Automated Measurement of Waveform Parameters GO NO GO Determination P...

Страница 5: ...C key the channel information appears If you press the ESC key again the channel information is cleared and the waveform display area is expanded horizontally SHIFT ZOOM key Sections 8 5 and 11 4 Disp...

Страница 6: ...SAVE key Sections 13 11 13 12 and 16 3 Stores the screen image data to the storage medium Pressing the SHIFT key followed by the IMAGE SAVE key displays a menu related to the saving of the screen imag...

Страница 7: ...h Section 7 2 Acquisition mode Section 7 3 Sample rate Section 5 2 T div Section 5 2 Trigger position Section 6 2 Display record length Appendix 1 Trigger level Chapter 6 Ground level Chapter 6 Channe...

Страница 8: ...ecord length of the Z2 zoom waveform display Record length of the normal waveform Z2 zoom range Screen When Using the Dual Capture Function Section 7 6 Using the Dual Capture Function in the User s mM...

Страница 9: ...ote Menu type on which the item is selected from the displayed selections The selected item switches each time the soft key is pressed The selected item is displayed highlighted Menu type on which a s...

Страница 10: ...ng to the selected item appears Sub menu For items lower in the hierarchy than the top menu a sub menu is displayed Moving the pointer to this position causes the pointer s appearance to change from t...

Страница 11: ...igger A trigger is activated when the specified timeout time is reached Period trigger A trigger is activated when the period of the signal matches the specified condition Window trigger A trigger is...

Страница 12: ...e Function in the User s Manual Part 1 This function enables trend recording in low speed roll mode1 while acquiring data at high speeds The function is useful when capturing at high speeds abnormal p...

Страница 13: ...24000 data points for all waveforms can be saved Of those up to 24 arbitrary parameters can be displayed on the screen Delay parameter 1 Time difference from the reference waveform or trigger point t...

Страница 14: ...n a specific determination is made include printing saving of the screen image data saving of the waveform data sounding of a beep and sending of an e mail message with the C10 option Zone determinati...

Страница 15: ...data such as saving the data to a file performing cursor measurements automated measurement of waveform parameters normal statistical processing and search zoom T Y Waveform Recording T Y waveforms ar...

Страница 16: ...Comparatively simple computation Addition subtraction multiplication division digital filter differentiation and integration Comparison of Normal Computation Math1 to Math8 and Computation on DSP Chan...

Страница 17: ...o use the DL750 DL750P in a safe manner read the warnings given in section 3 3 Installing Input Modules in the user s manual before installing the module Install the module in the input module install...

Страница 18: ...the warnings given in section 3 6 Connecting Probes and the cautions given in section 3 7 Compensating the Probe Phase Correction in the user s manual before connecting the probe Probe compensation ou...

Страница 19: ...Initialize soft key are as follows Date time settings language settings communication settings and SCSI ID setting To initialize all settings except the date time carry out the following procedure Tu...

Страница 20: ...he User s Manual Part 1 By default the screen is divided into 4 windows Quad Since only CH1 is observed in this case the display format is set to Single 2Display the selection menu and select Single 1...

Страница 21: ...g from 200 s div to 100 ms div Section 5 2 Setting T div in the User s Manual Part 1 Time axis setting refers to setting of the time per division of the grid If the time axis setting is increased the...

Страница 22: ...s div Section 5 2 Setting T div in the User s Manual Part 1 The display returns from roll mode to update mode and 5 periods of the waveform are displayed 1Turn the TIME DIV knob clockwise and set the...

Страница 23: ...ection of the specified trigger slope If auto setup is performed the trigger mode is set to auto For details see section 6 1 Setting the Trigger Mode in the user s manual Trigger position Determines t...

Страница 24: ...rm moves 2 div to the left If you wish to select 10 50 or 90 you can press the corresponding soft key Trigger position mark Pre trigger section Post trigger section Trigger point Change the Trigger Mo...

Страница 25: ...difference between the X axis values at Cursor1 and Cursor2 Y axis value of Cursor1 Y axis value of Cursor2 The difference between the Y axis values at Cursor1 and Cursor2 The CH specified by trace is...

Страница 26: ...1Press the ZOOM key 3Set the jog shuttle control to Z1 Mag 4Turn the jog shuttle to set the zoom ratio of Z1 5Likewise set Z2 Mag If the jog shuttle control is set to both Z1 Mag and Z2 Mag the zoom r...

Страница 27: ...P in the user s manual 1Press the PRINT key Printing is executed Output example Note Pressing the SHIFT key followed by the PRINT key displays the PRINT menu Though not required in the procedures give...

Страница 28: ...ta to a Specified Storage Medium Section 13 7 Saving Loading the Waveform Data in the User s Manual Part 2 Saves the data of the waveform displayed on the screen to the storage medium When the save op...

Страница 29: ...factory default Cancel initialization Applicable Waveforms for Auto Setup Auto setup can be performed on the following waveform Frequency Approx 50 Hz to 1 MHz Input voltage magnitude Up to approx 20...

Страница 30: ...abbreviated name NONISO_10M12 High Voltage 100 kS s 16 Bit Isolation Module with RMS MODEL 701260 abbreviated name HV with RMS Universal Voltage Temp Module Only When Measuring Voltage MODEL 701261 a...

Страница 31: ...to eight characters Unit Section 5 16 You can set the temperature unit to C K or F RJC Section 5 16 Select whether to perform reference junction compensation by the internal RJC circuit ON OFF Burnout...

Страница 32: ...Shunt Cal MODEL 701271 abbreviated name STRAIN_DSUB Bridge Voltage 2 V When the bridge resistance is 120 to 1000 5 V When the bridge resistance is 350 to 1000 10 V When the bridge resistance is 350 to...

Страница 33: ...5 17 Set the sensitivity of the acceleration sensor in the range of 0 10 to 2000 00 mV Unit Unit Section 5 12 Enter the acceleration unit The default setting is m s2 Waveform label Section 8 10 You c...

Страница 34: ...waveforms are zoomed expanded vertically and the offset voltage Note Turn ON OFF linear scaling Section 5 11 Set the method of scaling the measured values Select OFF when not scaling Waveform label Se...

Страница 35: ...ON in order from the top Auto Differences in the Bit Display Position Example when bit 7 is OFF Note or A1 A3 A2 A4 A5 A6 A8 A1 A3 A2 A4 A5 A6 A8 Fixed Auto Turn ON OFF the event waveform display Sect...

Страница 36: ...expanded vertically DSP Channel G3 Option DSP Digital Signal Processor channels are used to perform computations between channels and filter computations in realtime using the output data of the input...

Страница 37: ...ined at every time interval from the data sampled at the maximum sample rate of each input module The time interval used to determine the values is equal to the sample rate of the normal mode The maxi...

Страница 38: ...set to edge trigger meets the condition B Time B Time B TimeOut A trigger is activated on the falling or rising edge of the pulse when the pulse width width over which condition B is met exceeds or dr...

Страница 39: ...splay is updated once when the trigger conditions are met and the waveform acquisition stops Single N mode Section 7 5 When the trigger conditions are met waveform acquisition is repeated the specifie...

Страница 40: ...menu Buzzer Sounds a buzzer Send Mail Sends an e mail message to a specified address C10 option Note Action on trigger Section 6 18 Performing a specified action each time a trigger occurs Action on...

Страница 41: ...er position to the 50 position of the display record length To the 90 position Set the trigger position to the 90 position of the display record length Trigger Position Note Display record length Trig...

Страница 42: ...ON OFF Conditions That Allow Dual Capture Dual capture is possible when all of the following conditions are met Time div of the main waveform 100 ms div to 3 day div roll mode display Sample rate of...

Страница 43: ...d automated measurement values of waveform parameters Interpolation When the Area Is an Interpolation Area Interpolation When the Area Is not an Interpolation Area If the number of data points is grea...

Страница 44: ...axis range 5 to 5 div of the T Y waveform display Notes When Displaying X Y Waveforms The divided windows of the T Y waveform display when using the T Y X Y mode are displayed according to Format in...

Страница 45: ...Time Axis Parameters Display Example of Automated Measurement Note Target waveform for distal mesial and proximal Select the waveform used to determine the values of distal mesial and proximal that a...

Страница 46: ...eform cycle within the specified measurement range Statistical processing of history data Performs automated measurement of waveform parameters on the waveform that is acquired using the history memor...

Страница 47: ...limits of the automated measurement of waveform parameters GO NO GO determination is performed on whether the measured value enters the range or exits the range Zone GO NO GO is determined by creatin...

Страница 48: ...ursors and V cursors are displayed simultaneously Marker cursors Four markers are displayed on the specified waveform The voltage and the time from the trigger position of each marker as well as the v...

Страница 49: ...ified waveform parameter exceeded or stayed within a specified range History search conditions Section 11 2 When search mode is Zone Set the search zone search source channel search condition and sear...

Страница 50: ...computed data into the specified number of data points Cycle Count This is done across multiple periods of data from the start to the end position of the computation Then the average of the data poin...

Страница 51: ...to the newest position of the waveform Moves the zoom box of Z1 or Z2 to the newest position right end of the main waveform This is useful when displaying the newest waveform expanded in roll mode Sel...

Страница 52: ...xecute the search Edge Search Example when the edge is set to rising and the count is set to 2 Note The following figure shows the menu that appears when the search type is set to Edge Select the sear...

Страница 53: ...visions Display of past search results Specify the number of the search result to be displayed Execute the search Playing the Voice Memo That Is Found When using the voice search function the voice me...

Страница 54: ...tch on the earphone microphone is held down or from the time when the Rec Start soft key is pressed until the Rec Stop soft key is pressed Recording is possible up to a total of 100 s Fixed Recording...

Страница 55: ...tion 12 3 Select the print destination of the screen image from built in printer USB printer network printer and PDF DL750P only Output format Select the output format of the USB printer from ESC P ES...

Страница 56: ...en Image Drive Name of the Storage Medium FD ZIP PC Card HD HD 1 SCSI5 SCSI5 1 NetWork USB Note Floppy disk Zip disk DL750 only PC card Hard disk Partition 1 on the hard disk SCSI device with the ID n...

Страница 57: ...of the measurement Measurement and recording stops when the maximum number of divisions of measured values that can be saved is recorded Single Starts recording when the trigger conditions are met af...

Страница 58: ...elect whether to automatically record print to the built in printer when waveform acquisition is stopped Reprint Section 9 7 Displays the menu that recalls data recorded immediately up to the point wh...

Страница 59: ...e data and setup data to the FTP server on the network Access the DL750 DL750P from an FTP client on the network and retrieve the files on the storage medium of the DL750 DL750P the storage medium of...

Страница 60: ...IM 701210 07E 59 MEMO...

Страница 61: ...IM 701210 07E 60 MEMO...

Отзывы: