IM 12D03D02-01E
7-4
7
.
Measuring Principles of this Instrument
• Checking if meter is working correctly
When the temperature coefficient already set is accurate, the conductivity displayed
should be constant regardless of liquid temperature (within permitted temperature
range).
Check that the temperature coefficient already set is accurate. When the liquid
temperature is lowered, if a larger conductivity is indicated, the temperature
coefficient already set is too small; whereas, if a smaller conductivity is indicated,
the temperature coefficient already set is too large. In such a case, change the
temperature coefficient such that the measured conductivity does not change with
temperature.
7.4 Wetted Part Materials of Sensors
• For general-purpose sensor
Titanium (sensor)
Polyphenylene sulfite resin, Polypropylene resin (insulated area and cover)
Fluoro rubber (O-ring)
• Sensor for high purity water measurement
SUS 316 stainless steel (electrode element)
Polypropylene resin (insulated area)
Fluoro rubber (O-ring)
• Chemical-resistant sensor
Glass, platinum black (electrode element)
• Sensor for high-conductivity measurement
Glass, platinum black (electrode element)