Measurement
3. Measurement
3.1 Precautions
(1) Be sure to check that the cell constant and the temperature coefficient are set
correctly.
(2) Check that the plastic cover (for general-purpose sensor) and outer electrode
(for high purity water measurement) are secure.
(3) Do not use the SC72 meter to measure liquids with temperatures over 80°C.
(If the sensor grip is immersed, liquid temperature shall be below 50°C.) Do
not use the meter to measure extremely corrosive liquids such as solutions of
hydrofluoric acid.
(4) Remove dirt and stains from the meter with soft tissue. If necessary, wipe the
meter case with neutral detergent.
(5) If any problem with the meter arises during measurement, refer to the
Troubleshooting section later in this manual to determine the cause.
(6) After finishing measurement, flush stains on the sensor and measured solutions
with water, and store the meter (refer to Chapter 5, “Maintenance”).
When the meter is used on a table:
The meter is designed as a portable instrument; however, to use it on a table, attach
non-slip pads (supplied with the instrument) at top and bottom of the meter to stop it
from moving when the sensor is moved.
SC METER
PERSONAL
2004.
SC72
S1.0
E000001
MODEL
No.
STYLE
Non-slip pads
Figure 3.1 Position of Non-slip Seats