IM 12D03D02-01E
5-6
5
.
Maintenance
Procedures for calibration with standard solution
Whole measurement value
is flashing
Complete the calibration
Least significant figit of the
measuremet value is flashing
Select
CAL mode
POWER
F/ENT
F/ENT
F/ENT
F/ENT
Std. sol.
Water
Wash sensor with
water completely.
Wait until the measured
value is stable.
Set the conductivity of
the standard solution.
Thoroughly wipe off the
washing water from the
electrodes before immersing
the sensor in standard solution.
Refer to Section 4.3, “Function Modes,” for temperature coefficient and cell constant
settings.