IM 12D03D02-01E
2-4
2
.
Preparation
2.5 Setting Cell Constant
Even for sensors of the same type, each sensor has its own distinct cell constant. So,
set the proper cell constant as indicated on the sensor cable.
Whenever sensors are replaced, be sure to change the cell constant setting in the
meter accordingly. Cell constants once set are stored in non-volatile memory and are
not lost even when the batteries are replaced.
l
Setting cell constants
Press the
F/ENT
key to switch to function mode. Then select the C.C display with the
keys and access the cell constant setting display with the
F/ENT
key. Use the
keys to set the cell constant, then press the
F/ENT
key to confirm it.
Refer to Sec. 4.3 (4).
F/ENT
F/ENT
F/ENT
MEAS
for Function mode
for C.C panel
to enter C.C
select C.C
You can abort this procedure at any time by pressing
MEAS
to revert to
measurement mode.