IM 12D03D02-01E
6-1
6
.
Troubleshooting
6. Troubleshooting
6.1 Causes of Abnormal Conductivity Display
The measured temperature is stable, but the conductivity reading is unstable, or an
unrealistic value is displayed, check the following:
(1) Is maintenance required, and has handling of the meter been adequate?
(2) Does the battery need replacement?
(3) Is SC72 meter unserviceable?
If any problem is evident, follow the procedure shown in Section 6.2 to determine the
cause and fix the problem. If you cannot fix it, contact your nearest Yokogawa sales
office.
Abnormal
Components need replacement
• Displayed value unstable
• Displayed value unrealistic
• Error message displayed
• Battery
Needs maintenance
• Needs recalibration with standard solution
• Electrode stained
• Electrical leakage in connector
SC meter failure
• Contact Yokogawa.
Figure 6.1 Causes of abnormality
6.2 Error Messages and Corrective Action
Table 6.1 Error Message
Error Message *1
Description
Err1
Temperature compensation computation value out of range
Err2
Out of temperature measurement range
Err3
Calibration error
Err6
Meter electronics failure
or
Out of measuring range
*1: For displayed digital characters, see Alphanumeric Display Table in Preface.