<7. EXPLOSION-PROTECTED TYPE INSTRUMENTS>
7-9
IM 01R01B30-00E-E 8th edition November 01, 2019-00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
U
Supply 230 V AC or 115 V AC or 24 V DC
Limit switch MAX
Rotameter RAKD -T80
Limit switches
Supply 230 V AC or 115 V AC or 24 V DC
-
+
-
+
+
-
BU
2
BN
BU
BN
1
2
1
MIN
F
MAX
S
T
KEMA 00ATEX1037X
SJ2-SN
Limit switch, option /K6, /K7 or /K8
If only one limit switch is used, power supply option /W2E (KHA6-SH-Ex1)
for 230/115V AC or power supply option /W4E (KFD2-SH-Ex1) for 24 V DC
can be used.
Limit switch power supply, option /W2F or /W4F
2 x KHA6-SH-Ex1(230/115 V AC)
2 x KFD2-SH-Ex1(24 V DC)
PTB 00ATEX 2043 (230/115 V AC)
PTB 00ATEX 2042 (24 V DC)
~
~+
-
EN 60947-5-6 (Namur)
23
24
24
23
~
Hazardous area
Safe area
Limit switch MIN
-
+
12
10
12
10
U
~
~ +
-
~
T, S: QUICKON connector
C = 30 nF
Ex ia IIC T6...T4 Gb
L = 100 µH
P = 64 mW
I = 25 mA
U = 16 V
i
i
i
i
i
C = 3.6 µF
I = 16.8 mA
P = 41 mW
U = 9.56 V
L = 130 mH
[Ex ia Ga] IIC
o
o
o
o
o
Fig. 7-6 RAKD with ATEX approval (option /KS1) with fail-safe limit switches and AC/DC power
supply