<3. INSTALLATION>
3-1
All Rights Reserved. Copyright © 2003, Rota Yokogawa
IM 01R01B30-00E-E 8th edition November 01, 2019-00
3. Installation
3.1 Installation in the pipeline
The pipe has to be stabilized so that vibrations on the pipe are avoided.
If contamination or soiling of the RAKD is to be expected, a bypass should be installed to allow the removal of
the instrument without interruption of the flow. (see. chapter 6.1 "maintenance").
Before installing the Rotameter, ensure that no parts of packing, securing or residues of the sealing compound
get inside of the Rotameter.
To avoid stress in the connecting pipes, the connecting flanges must be aligned in parallel and axial direction.
Bolts and gaskets have to be selected according to the maximum operating pressure, the temperature range
and corrosion conditions. Gaskets need to be positioned centered and the nuts should be tightened with a torque
appropriate for the pressure range.
NOTE
The recess in front of the spanner flat shall not be used to tighten the process connections.
Please read also chapter 2.2. For further instructions on installation please refer to VDI/VDE3513.
In case of devices without valve but with process connection internal thread, the float shaft of tubes with cone
44 and bigger sizes may move into the range of the threaded joint. Keep attention not to bend the guide bar
when tightening the screw connection the connection. The inner diameter of the connector must be at least
8 mm for cone sizes 44 to 51 resp. 10 mm for cone sizes 52 and 53.
To avoid float bouncing in case of gas applications, attend to VDI/VDE 3515 sheet 3.
For devices without valve use a throttle either in the inlet or outlet (to install at that side with the bigger volume).
For installation of several instruments in parallel tubes ensure that the distance between the middle axis is
at least 120 mm to avoid magnetic influence. The distance to other ferritic material should not be less than
60 mm. Ensure that the strength of external magnetic fields is approximately 0 mT.
3.2 Wiring
3.2.1 Connecting diagrams
Please consider the drawings of this chapter.
RAKD with electronic transmitter or with limit switches have one or two Quickon connectors at the rear.
In the connecting diagrams the lower Quickon connector is named "S" and the upper one "T".
Connector positions wich are not in use are closed with a blind plug.
With option /A29 or /A30 M12 connectors are mounted, see 3.2.3.
The installation drawings of this chapter show Quickon connectors.
Tab. 3-1 Connections for the concerning configurations
Type T
without limit
switch
Type T
with MIN
limit switch
/K1 or /K6
Type T
with MAX
limit switch
/K2 or /K7
Type T
with MIN/
MAX
limit switch
/K3 or /K8
Type E
without limit
switch
without puls
output
Type E
with MIN
limit switch
/K1 or /K6
Type E
with MAX
limit switch
/K2 or /K7
Type E
with puls
output
/CP
Quickon
above "T"
---
---
MAX limit
switch
MAX limit
switch
Analog
output
Analog
output
Analog
output
Analog
output
Quickon
below "S"
---
MIN limit
switch
---
MIN limit
switch
---
MIN limit
switch
MAX limit
switch
Pulse output
The load resistance of metering or indicating instruments, which are connected serial to the current output, may
not exceed (U - 14 V) / 20 mA .