<5. Converter Operation>
5-31
IM 12F4A1-01E
(b) Every time [>] key is pressed, _ _ 1 and _ _ 2 are
alternately displayed.
→ _ _ 1 ↔ _ _2
(c) Select _ _ 1 (mg/l) or _ _ 2 (ppm).
→ _ _ 2
(d) Press [ENT] key to enter it.
→ _ _ 2
(Flashes once.)
The selected unit appears to the right of data when a residual chlorine concentration is displayed
(see Figure 5.12).
MEAS.
MAINT.
PROGRAM. 1
PROGRAM. 2
ppm
mg/l
Residual chlorine concentration display
Either one of the units set
with function code “c.”is
indicated.
F5.12.ai
Figure 5.12
Residual Chlorine Concentration Unit Indication
(19) Selection of Air Pump Action Function “execution/stop”
(Function code “d.”)
An air pump for air purge is added only when it is specified.
The air pump action function is that which always places the air pump in operating status in
<MEAS.> mode.
Hence, if “on (execution)” is set with this function code, the air pump automatically enters the
operating status even if the mode is transferred to <MEAS.> mode with the air pump action
stopped.
If the air pump is added, it has been set at “on” on shipment.
Thus it is unnecessary to specifically set it.
If it is to be changed to “oFF (stop)”, set it again by the following key operation:
(a) Select function code "d.".
→ on
(Execution)
oFF
(Stop)
(b) Every time [>] key is pressed, on and oFF are alternately
displayed.
→ on ↔ oFF
(c) Select on (execution) or oFF (stop).
→ oFF
(d) Press [ENT] key to enter it.
→ oFF
(Flashes once.)
oFF
(Cleaning stop)
(b) Every time [>] key is pressed, on and oFF are alternately
displayed.
→ on ↔ oFF
(c) Select on (execution) or oFF (stop).
→ oFF
(d) Press [ENT] key to enter it.
→ oFF
(Flashes once.)
(20) Selection of negative (below-zero) display in <MEAS.> mode
(Function code “E.”)
When this function is set to on (enabled), a negative value (below 0.00 mg/l) is not displayed
even if the measured residual chlorine concentration value is negative. However, this setting can
only be enabled in <MEAS.> mode.
The factory setting is disabled. If the setting is to be changed, perform the following key
operation.
(a) Select function code “E.”.
→ on
(enabled)
oFF
(disabled)
(b) Every time [>] key is pressed, on and oFF are alternately
displayed.
→ on ↔ oFF
(c) Select on (enabled) or oFF (disabled)
→ on
(d) Press [ENT] key to enter it.
→ on
(Flashes once.)