8.10.1
Check/Maintenance/Replacement of Rotating contact
CAUTION
Never touch Rotating contact except when replacement is performed.
When you hold Rotating contact, pinch it with two fingers and never touch the center contact part.
Never drop it or gives a shock.
Contact part
CAUTION
The storage limit for Rotating contact is 1 year after the purchase, considering degradation of the
lubricant used inside. Store it in room temperature and keep it away from direct sunlight.
[Checks and Maintenance]
Rotating contact does not need inspection. Never contact the rotating contact except when the
replacement is performed.
[Replacing the Rotating contact]
If they are replaced by the customer, follow the steps below.
(1) Shut off the power fed to the free available chlorine analyzer to stop the operation.
Gear head
Terminal block for
Rotating contact
Motor assembly
Drive shaft
Driven shaft
Drive belt
Pulleys
Clamp
Rotating contact
Plate
Motor
adapter plate
Indicator electrode
Counter electrode
Electric holder
Base
Figure 8.7
Electrode Mechanism Block