<4. Operation>
4-5
IM 12F4A1-01E
Table 4.3
Reagent Composition and Its Component Quantities
Measurement
Object
and Max.
Concentration
Residual Chlorine
Required quantity of containers for each
measurement (approx. for 45 days)
Up to 6 mg/l
More than 6 mg/l to
10 mg/l (inclusive)
Quantity in 100 l
Up to 6 mg/l
More than 6 mg/l to
10 mg/l (inclusive)
KI
500 g
1000 g
500 g x 1
500 g x 2
CH
3
COONa
150 g
500 g x 1 (3 times)
CH
3
COOH
1000 ml
500 ml x 2
4.2.2
Filling Bead Case with Electrode Polishing Glass
Beads
Contamination adhering to the gold alloy surface of the gold alloy rotating electrode results in
measured value error. To prevent such error owing to contamination, the gold alloy surface is
continuously polished with glass beads during operation.
Fill the bead case in the measuring tank with glass beads which are delivered by placing in the
accessory box.
n
Glass Beads Filling Procedure
(1) Remove the electrode mechanism block by loosening two fixing screws.
(2) Put the glass beads into the bead case.
Feed the glass beads up to the lowest slit position (approx. 4 cm above the bottom) of the
bead case.
(3) Restore the electrode mechanism block.
Note: The upper surface of the glass beads with the indicator electrode placed in the bead case is to be located 1 cm higher than the
lowest slit position.
F 4.2.2.ai
Bead case
Counter electrode
Bead case cover
Lowest slit
position
Measuring tank
Fixing screws
(2 pcs)
Indicator electrode
Glass beads
Status in which the electrode
mechanism block has been
removed.
Electrode mechanism
block
Figure 4.4
Detector Construction
14th Edition : Jul. 15, 2020