4.3.10
Polishing Indicator Electrode
The gold alloy surface of the indicator electrode must be kept clean. Polish the gold alloy surface
in the following procedure:
(1) Confirm that the mode is <MAINT.> and then press [CELL] key on the control panel to
stop revolution of the indicator electrode. When the electrode stops rotating, <CELL> lamp
extinguishes.
(2) Remove the electrode mechanism block from the measuring tank and set it as shown in
Figure 4.7. The electrode mechanism block is fixed with two screws to the measuring tank
bracket.
Electrode mechanism
block
Fixing screws
F4.6.ai
F 4.7.ai
Electrode mechanism
block cover
Measuring tank
Figure 4.6 Electrode mechanism block fixing screws Figure 4.7 Measuring tank status in maintenance
(3) Remove the indicator electrode.
Remove the electrode mechanism block cover of the detector. Detach the rotating electrode
with the end of the driven shaft supported by hand so that the driven shaft does not rotate.
Since the electrode is screws into the driven shaft, turn the electrode counterclockwise.
Do not to touch the rotating contact.
Driven shaft
Rotating contact
Indicator electrode
Collar
Gold alloy surface
Figure 4.8
Appearance of electrode mechanism block and Indicator electrode