<2. Specifications>
2-2
IM 12F4A1-01E
Sample conditions:
Temperature:
0 to 50 °C
pH:
3 to 9 pH
Flow:
1 to 4 l/min, for pure water or municipal water application.
5 to 10 l/min, for raw water or sea water application.
10 to 20 l/min, for secondary sewage treatment application.
Pressure:
20 to 500 kPa
Electrodes:
Indicator electrode:
Rotating gold alloy electrode
Counter electrode:
Platinum electrode (built-in Pt1000 Ω RTD), the combined chlorine
insensitive version uses a silver chlorine electrode (built-in Pt1000
Ω RTD).
Reagents:
Composition:
The following reagents are dissolved in pure water to give 100 l of
solution.
(Note) Pure water should be equal to or exceed level of tap water and should not contain oxidizing and reducing substances
such as chlorine.
For residual chlorine
measurement
For free chlorine
measurement
For combined chlorine
measurement
Maximum
concentration
Up to 6 mg/l Over 6 up to 10 mg/l
Up to 10 mg/l
Up to 10 mg/l
Potassium iodide,
extra pure
500 g
1000 g
---
---
Potassium bromide,
extra pure
---
---
4000 g
4000 g
S o d i u m a c e t a t e ,
extra pure
150 g
1000 g
5400 g
Acetic acid, extra
pure
1000 ml
1000 ml
200 ml
Consumption:
1.5 ml/
min ±10%
Constant flow pump:
Sample flow:
50 m
l/
min ±10%
Reagent flow:
1.5 m
l/
min ±10%
Converter functions:
Display functions:
Data:
Concentration, temperature, applied voltage, diffusion current,
output signal %, pump flow, zero point, slope
Status indicators: Measurement, maintenance, hold, calibration in progress,
cleaning, fail
Operating status : Rotating gold alloy electrode, mertering pump, air pump,
solenoid valve ON/OFF
Diagnostic functions: Concentration over range, measurement temperature error, no liquid in
flow cell, converter error, temperature compensation over range, setting
upper limit exceeded, zero point error, slope error, response error
Maintenance functions (MAINT mode):
One-touch calibration, flow cell and sand filter system cleaning, pump
flow measurement, one-touch plateau characteristic acquisition, failure
error information
Setup functions:
PROGRAM 1 mode
Output Range 1 and Range 2 settings, 2-line-segment output setting,
applied voltage setting, response time and stability parameter settings,
contact output during FAIL condition (ON/OFF), indication hold during
MAINT (ON/OFF), remote range switching ON/OFF