<5. Converter Operation>
5-6
IM 12F4A1-01E
5.2.4
Function Codes and Their Description in
<PROGRAM.2> Mode
Table 5.6 shows the types of function code and their description in <PROGRAM.2> mode.
Table 5.6
Functions List in <PROGRAM.2> Mode
F
Description
Setting Range
Display
(example)
Initial
Value
Remarks
1
2
3 Setting measuring tank cleaning cycle
0.1 to 24.0 hours
0001.0 hours
1.0
Applied to RC400G-3
analyzer only
4 Setting measuring tank cleaning time
0.1 to 25.0 min
0001.0 min
1.0
5 Setting relaxation time after cleaning
0.1 to 25.0 min
0005.0 min
5.0 (3.0)
The value in ( ) is for
RC400G-3 analyzer.
6
Setting and filter cylinder cleaning
cycle
0.1 to 24.0 hours
0002.0 hours
2.0 (0.5)
The value in ( ) is for
RC400G-3 analyzer.
7
Setting sand filter cylinder cleaning
time
0.1 to 25.0 min
0001.0 min
1.0
8 Setting substitution time
0.1 to 25.0 min
0006.0 min
6.0
Applied to RC400G-3
analyzer only
9 Setting cross time
0.1 to 25.0 min
0001.0 min
1.0
Applied to RC400G-3
analyzer only
A Setting automatic zero calibration
cycle
0.1 to 99.9 days
0010.0 days
10.0
b Setting automatic zero calibration
preparation time
1 to 100 min
0010 min
10
c Setting automatic zero calibration
relaxation time
1 to 100 min
0010 min
10
d
E
F
1.
2.
3.
4.
Temperature compensation range
exceeding detection function
"execution/stop"
on/oFF
on
5. Residual chlorine concentration upper
limit alarm function "execution/stop"
on/oFF
on
6.
7.
8.
9.
A.
b.
c.
d.
E.
F.
Note 1: If an unused function code is selected, "- - - - -" is displayed.
Note 2: If a value outside the setting range is set, "not" is displayed (pressing [>] key returns the display to data display).
Note 3: Values shown in the initial value column are those to be shown when the memory that stores data is initialized.
14th Edition : Jul. 15, 2020