8-14
<Toc> <Ind>
<8. PR201 Original Communication Protocol>
IM 77C01E01-10E
• Example
Reads the maximum voltage-1 of PR300 with station number 01.
䊐䊐
indicates the
checksum value.
[Command]
[STX]DGB01
䊐䊐
[ETX][CR]
Returns a response of the maximum voltage-1: 1000 V (1.000E+3 V) for the above
command.
[Response]
[STX]DGB011.000E+3
䊐䊐
[ETX][CR]
●
Parameter: D (Reads maximum current-1), Parameter: E (Reads maximum
current-2. Old style), Parameter: S (Reads maximum current-2), Parameter: T
(Reads maximum current-3)
• Function
Reads the maximum current-1.
Reads the maximum current-1 by parameter: D, the maximum current-2 by parameters: E
and S, and the maximum current-3 by parameter: T.
• Command/response
Number
of bytes
1
2
1
2
2
1
1
Command
element
STX
DG
D
(E)
(S)
(T)
Station
number
(ADR)
Checksum ETX
CR
There is no command data for data reading.
1
2
1
2
8
Response
element
STX
DG
D
(E)
(S)
(T)
Station
number
(ADR)
ⵧ
.
ⵧⵧⵧ
E
䊊
2
1
1
Checksum ETX
CR
Number
of bytes
Response data is returned as an 8-byte ASCII character string (0.000E+0 to 9.999E+5).
• Example
Reads the maximum current-1 of PR300 with station number 01.
䊐䊐
indicates the
checksum value.
[Command]
[STTX] DGD01
䊐䊐
[ETX][CR]
Returns a response of the maximum current-1: 1000 A (1.000E+3 A) for the above
command.
[Response]
[STX]DGD011.000E+3
䊐䊐
[ETX][CR]
Содержание PR300