Standard Protocol
Description Parameter p4, the terminal block serial number,
is undefined if the terminal block unit is not
connected.
IT
Outputs terminal block status
(query only).
Syntax
IT p1<terminator>
p1
Terminal block status
When the 8-ch terminal block is connected
0
When the 16-ch terminal block is connected
1
When the terminal block is not connected
3
Query
IT?
5.9
Output Commands (Setting/
Measurement/Data Output)
FC
Outputs the screen image data
Syntax
FC p1<terminator>
p1
Screen image data output
GET
Example
Output the screen image data from the XL100.
FC GET
Description Captures the current displayed screen on the
XL100 and outputs the data in BMP format.
FE
Outputs the setting data
Syntax
FE p1,p2,p3,p4,p5,p6,p7,p8,p9,p10,
p11<terminator>
p1
Output data type
Setting data of setting commands:
0
Decimal place and unit information:
1
Setting data of basic setting commands:
2
Reserved:
3
Reserved:
4
p2
First channel number
For 8 ch input:
01
to
08
,
For 16 ch input:
01
to
16
<Numeric value>
00
(OFF)
p3
Last channel number
For 8 ch input:
01
to
08
,
For 16 ch input:
01
to
16
<Numeric value>
00
(OFF)
p4
First pulse measurement channel number:
PL01
p5
Last pulse measurement channel number:
PL01
p6
First logic input channel number
DI01
to
DI02
p7
Last logic input channel number
DI01
to
DI02
p8
First calculation channel number
CA01
to
CA32
p9
Last calculation channel number
CA01
to
CA32
p10
First communication channel number
CO01
to
CO32
p11
Last communication channel number
CO01
to
CO32
Example
Get the decimal place and unit for channels 01 to
12 and pulse channel.
FE 1,01,12,PL01,PL01
Description • Set the first channel number and last channel
number parameters so that the last channel
number is greater than or equal to the first
channel number.
• Parameters p2, p3, p4, p5, p6, p7, p8, p9,
p10, and p11 are valid when p1 is set to 0, 1,
or 2.
5.8 Output Commands (Control)/5.9 Output Commands (Setting/Measurement/Data Output)