5-26
IM XL100C-E
• For parameters p5 and p6, enter a value using
5 digits or less excluding the decimal point.
The decimal position is fixed to the position
indicated in the tables in section 5.4, “Input
Range Parameters.”
• For parameters p7, p8, and p9, either set all
three parameters or omit all three parameters
when logging is stopped.
• Set parameters p10 and p11 according to the
tables in section 5.4, “Input Range
Parameters” when p11 is not ANY.
SO
Sets the calculating equation
When setting the channel to OFF (SKIP)
Syntax
SO p1,p2<terminator>
p1
Calculation channel number
CA01
to
CA32
31
to
42
(*)
p2
Calculation ON/OFF
ON
,
OFF
Query
SO[ p1]?
Example
Set calculation channel 01 to OFF.
SO 01,OFF
Description • This command cannot be specified while
logging is in progress.
• Calculation channels that are set to OFF are
not calculated.
• Setting parameter p1 to 31 to 42 is equivalent
to specifying CA21 to CA32.
When setting the channels to ON
Syntax
SO p1,p2,p3,p4,p5,p6,p7,p8
<terminator>
p1
Calculation channel number
CA01
to
CA32
31
to
42
(*)
p2
Calculation ON/OFF
ON
,
OFF
p3
Calculating equation (up to 40 characters,
differential calculation only)
p4
Span lower limit
-9999999
to
99999999
p5
Span upper limit
-9999999
to
99999999
p6
Span decimal place
x.xxxx:
0
xx.xxx:
1
xxx.xx:
2
xxxx.x:
3
xxxxx:
4
p7
Unit (up to 6 characters)
p8
Unit group
Sets the unit group (see section 5.4, “Input
Range Parameters”).
Query
SO[ p1]?
Example
Set the calculation channel to CA01, the
calculation to ON, the computing equation to “the
difference between channel 01 and 02”, the span
lower limit to –10.000, the span upper limit to
15,000, and the unit to V.
SO CA01,ON,01-02,-10000,15000,4,V,
VOLT.CURR.
Description • Parameters p4 and p5 can be set only as
display span settings while logging.
• For parameters p4 and p5, enter a value using
5 digits or less, excluding the decimal, for
negative numbers (excluding the sign) and
positive numbers.
• For parameters p4, p5, and p6, either set all
three parameters or omit all three parameters
when logging is stopped.
• Setting parameter p1 to 31 to 42 is equivalent
to specifying CA21 to CA32.
• The parameters that can be used in the
equation of p3 are as follows.
For 8 ch input:
CH01 to CH08 or 01 to 08
For 16 ch input:
CH01 to CH16 or 01 to 16
Calc. CH:
CA01 to CA32
Comm. CH:
CO01 to CO32
Constant:
Within the following ranges
with fixed decimal place
-9.9999 to 9.9999
-99.999 to 99.999
-999.99 to 999.99
-9999.9 to 9999.9
-99999L to 99999L
(Add “L” for integers.)
• Set parameters p7 and p8 according to the
tables in section 5.4, “Input Range
Parameters” when p8 is not ANY.
SM
Sets the communication channel
When setting the communication channel to OFF
(SKIP)
Syntax
SO p1,p2<terminator>
p1
Communication channel number
CO01
to
CO32
p2
Communication channel ON/OFF
SKIP
Query
SM[ p1[,p2]]?
Example
Set channel 01 to OFF (SKIP).
SM 01,SKIP
Description • This command cannot be specified while
logging is in progress.
• Measurements are not made on channels that
are set to OFF.
When using communication channels
Syntax
SO p1,p2,p3,p4,p5,p6,p7,p8
<terminator>
p1
Communication channel number
CO01
to
CO32
p2
Communication channel ON/OFF
ON
p3
Scaling value:
-1.0000E+38
to
-1.0000E-38, 0,
1.0000E-38
to
1.0000E+38
(The + sign of “E+” can be omitted.)
5.5 Setting Commands (Setting)