2.9
Setting the Web Server Function
To use the Web server function, set the following parameters in addition to those
described in section 2.3, “Setting the Ethernet Interface.”
• Enabling/Disabling the Web Server Function
Select ON (enable) or OFF (disable).
• Page Type (Type of Screen to Be Displayed)
• Monitor
• The screen displayed on the XL100 is displayed.
• The following information can be displayed.
• Alarm summary
• Measured and calculated data of all channels
• Logs (message log, error log, key login log, FTP log, e-mail log, and Web
operation log)
• For display examples, see section 2.10.
• Operator
All operations except key lock are possible in addition to the functions on the
monitor page.
• Monitor Page
• Enabling/Disabling the Monitor Page
• ON
The monitor page can be displayed on a Web browser.
• OFF
Disables the monitor page.
• Enabling/Disabling Access Control
• ON
Enables access control. You must enter the user name and password to display
the monitor page.
• OFF
Disables access control.
• Setting the User Name
Enter the user name using up to 16 alphanumeric characters using SYSTEM >
USER REGISTER.
• Setting the Password
Enter the password using up to six alphanumeric characters.
• Operator Page
• Enabling/Disabling the Operator Page
• ON
The operator page can be displayed on the browser.
• OFF
Disables the operator page.
• Enabling/Disabling Access Control
• ON
Enables access control. You must enter the user name and password to display
the operator page.
• OFF
Disables access control.